Diesel NOx Measurement Analyzer

Model : JDA-3000

  • Features

    Obtained type approval in Korea

    Pre-sampling device for measuring NO and NO2

    Pre-conditioning device to sample NO and NO2

    Pump pressure & flow display and data logging function

    Manageable maintenance and repair

    PM particle and moisture removal function

    NO - NDIR, NO2 - NDUV

Pre-conditioning Device

SUS Filter 10 micron Operating Temp. 100¡É

NO, NO2 Bench

Operating Temp 0 ~ 45¡É Operating Humidity 90% Max
NO Measuring Method NDIR (Non-dispersive infrared) NO Measuring Range 0 ~ 3,000 ppm
NO2 Measuring Method NDUV (Non-dispersive ultra violet) NO2 Measuring Range 0 ~ 2,000 ppm

Analyzer

Power Supply DC 12V ~ 24V Operating Humidity 90% Max
Operating Temp -10¡É ~ 45¡É Warm-up Time Within 20 min
Weight 15 kg Dimension 435Wx475Dx210H [mm]
Communication RS-232C